Xx.xx – JULABO SC 10000w Eco-Series Recirculating Coolers User Manual
Page 12

Operator responsibility – Safety instructions
12
Eco- Series
Temperature selection
digital
via keypad
indication on LED-Multi-Display
remote control via personal computer
indication on monitor
Temperature indication
°C
LED-Multi-Display
Resolution
°C
0.1
Electrical connection:
Computer interface RS232
Table 1
Pump capacity
Bath fluid:
Water, Silicone -oil
XX.XX
PUMP
1
XX.XX
PUMP
2
Circulating pump:
P3
P4 P3
P4
Flow rate max.
Lpm at 0 bar
30
40 33
43
Pressure max.
bar at 0 liter
1.8
2.5 3.5
4.4
Bath fluid:
Galden
®
eg.
Fluorinert
®
3283
Pressure max.
bar at 0 liter
3.5
4.4
Notice:
If Galden
®
or Fluorinert
®
is used the the charge of the motor increases. Wrong
adjustment causes overheating and eventually destruction of the motor.
With the circulation pump P3 as well as P4 a maximum pump pressure stage
>PUMP 1< may be adjusted.
The pump type can be recognized in the order no. on the name plate.
95 xx xxx _ xx PX xx