JULABO MA Heating Immersion Circulator User Manual
Page 43

MA
43
Error in A/D converter
Excess temperature protector defect.
The protection temperature is below the set working temperature
setpoint.
Set the protection temperature to a higher value.
The cable of the excess temperature protector has been disconnected or
short-circuited
The early warning system for low level reports a critical fluid level. Refill
bath fluid.
60
30
15
45
4 s
By quickly switching off and restarting the unit the alarm is
cancelled.
If the error occurs once more after the restart, a remote diagnosis is
required.
„Configuration Error“
The configuration of the circulator does not correspond with its current
application.
Press the
key for a non-recurring, automatic change of the
configuration.
In this case please call the JULABO Technical Service or an authorized
dealer.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic current
limiter. If viscosity of the bath fluid is or becomes too high, the motor
stops running.
15
A
M
P
Mains circuit breakers (resettable) 15 A
- AWC100 Compact Recirculating Cooler (15 pages)
- CF30 Cryo-Compact Circulators (35 pages)
- CF31 Cryo-Compact Circulators (76 pages)
- ED Heating Immersion Circulator (23 pages)
- ED-5A/B Bath Circulators (24 pages)
- ED-5A Open Bath Circulators (24 pages)
- EH Heating Immersion Circulator (27 pages)
- EH-5 Open Bath Circulator (28 pages)
- F250 Recirculating Cooler (35 pages)
- F12-ED Refrigerated / Heating Circulator (26 pages)
- F12-EH Refrigerated and Heating Circulators (32 pages)
- F12-MA Refrigerated and Heating Circulators (65 pages)
- F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath (69 pages)
- F38-EH Refrigerated and Heating Circulators (30 pages)
- F25-ME Refrigerated and Heating Circulators (66 pages)
- F25-HE Refrigerated and Heating Circulators (86 pages)
- F34-ED Refrigerated / Heating Circulator (25 pages)
- F25-HL Refrigerated Circulators (94 pages)
- FL7006 Recirculating Coolers (34 pages)
- FL2503 Recirculating Coolers (36 pages)
- FL20006 Recirculating Coolers (32 pages)
- FL1201 Recirculating Coolers (33 pages)
- FL300 Recirculating Coolers (48 pages)
- FK30-SL Calibration Baths (88 pages)
- FT200 Immersion Coolers (20 pages)
- FC600_600S Recirculating coolers (37 pages)
- FC1200S Recirculating cooler (45 pages)
- F81-HL Ultra-Low Refrigerated Circulator (93 pages)
- F70-ME Ultra-Low Refrigerated Circulators (64 pages)
- FP52-SL Ultra-Low Refrigerated Circulators (103 pages)
- FP51-SL Ultra-Low Refrigerated Circulators (90 pages)
- FP45-HE Refrigerated and Heating Circulators (79 pages)
- HL-4 Heating Circulators (89 pages)
- HE-4 Heating Circulators (78 pages)
- F81-HL Ultra-Low Refrigerated Circulators (93 pages)
- FT402 Immersion Coolers (24 pages)
- ME-4 Heating Circulators (62 pages)
- ME-31A Visco-Baths (62 pages)
- ME Heating Immersion Circulator (61 pages)
- MB-17 Open Heating Bath Circulators (52 pages)
- MB Heating Immersion Circulator (48 pages)
- Magnum 91 Temperature System (77 pages)
- MA-4 Heating Circulators (51 pages)
- LH 47 Highly Dynamic Temperature Control Systems (76 pages)