JULABO MA Heating Immersion Circulator User Manual
Page 18

Preparations
18
5.2.
Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly with
regard to the fire point!
If a bath fluid with a fire point of
65 °C is used, only supervised operation is
possible.
Water: The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for
temperature control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Recommended bath fluids:
Bath fluid
Temperature range
soft/decalcified water
5 °C to 80 °C
JULABO bath fluids
JULABO
Description
Thermal
G
Thermal
M
Thermal
HS
Order Number
10 liters
8 940 124
8 940 100
8 940 102
5 liters
8 940 125
8 940 101
8 940 103
Temperature range
°C
-30 ... 80
40 ... 170
20 ...250
Flash point
°C
--
284
270
Fire point
°C
--
306
>360
Color
light yellow
clear
light brown
JULABO
Description
Thermal
H10
Thermal
H20S
Thermal
H200
Order Number
10 liters
8 940 114
8 940 108
8 940 134
5 liters
8 940 115
8 940 109
8 940 135
Temperature range
°C
-20 ... 180
0 ... 220
60 ...200
Flash point
°C
190
230
292
Fire point
°C
216
274
334
Color
clear
light brown
clear
See website for list of recommended bath fluids.
ATTENTION:
The maximum permissible viscosity is 50 mm
2
/s