JULABO MA Heating Immersion Circulator User Manual
Page 22

Operating procedures
22
6. Operating
procedures
6.1. Power
connection
Caution:
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
The power supply plug serves as safe disconnecting device from the line
and must be always easily accessible.
Never operate equipment with damaged mains power cables.
Regularly check the mains power cables for material defects (e.g. for
cracks).
We disclaim all liability for damage caused by incorrect line voltages!
Check to make sure that the line voltage matches the supply
voltage specified on the identification plate.
6.2. Switching
on / Start – Stop
6.2.1. Switching
on the circulator
Switching on:
Turn on the mains power switch (1).
The unit performs a self-test. All segments of the 4-digit MULTI-
DISPLAY (LED) and all indicator lights will illuminate.
Then the software version (example: tt 2, V1.12, b004))
appears. The display „OFF“ or „R OFF“ indicates the unit is
ready to operate.
The circulator enters the operating mode activated before
switching the circulator off:
keypad control mode (manual operation)
or
remote control mode (operation via personal computer).
Start:
Press
key.
The actual bath temperature is displayed on the LED-
DISPLAY. The circulating pump starts with a slight delay.
Stop:
Press
key.
or
Keep
key pressed.
The LED -DISPLAY indicates the message "OFF".