2.4. Technical
specifications
Heating Immersion Circulator
MA
Working temperature range
°C
20 ... 200
Temperature stability
°C
±0,01
Overall dimensions (WxDxH)
cm
13x15x33
Usable bath depth
cm
from 8 to 14.5
Weight kg
4.0
All measurements have been carried out at: rated voltage and frequency
operating temperature: 70 °C ambient temperature: 20 °C bath fluid: water
Technical changes without prior notification reserved.
Mains power connection
208V - 230V; 50Hz - 60Hz
V; Hz
187 - 253; 50 - 60 (perm. range)
Current draw
(at 208V)
A
8
Current draw
(at 230V)
A
9
Mains power connection
100V - 115V; 50Hz - 60 Hz
V; Hz
90 - 127; 50 - 60 (perm. range)
Current draw
(at 100V)
A
8
Current draw
(at 115V)
A
9
MA
Temperature selection
digital
via keypad
indication on
MULTI-DISPLAY(LED)
remote control via personal computer
indication on monitor
Temperature indication
MULTI-DISPLAY (LED)
Resolution (-9.99 …. +99.99 = 0.01)
°C
0.01 / 0.1
Absolute Temperature Calibration °C
±3
Temperature control
PID
Heater wattage
(at 230 V)
kW
2,0
Heater wattage
(at 115V)
kW
1,0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate
max.at 0 bar
l/min
11 ... 16
Pressure max.
max. at 0 l
bar
0.23 ... 0.45
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
Computer
interface
RS232
Ambient temperature
°C
5 ... 40