Troubleshooting guide / error messages
42
10. Troubleshooting guide / error messages
+
Alarm with complete shutdown:
If one of the following failures occur a complete, all-pole shutdown of the 
heater and circulating pump is effected. 
„
“ lights up and a continuous signal sounds.
The code for the cause of alarm is indicated on the MULTI-DISPLAY 
(LED). 
Warning without a complete shutdown of the unit
The MULTI-DISPLAY (LED) indicates the cause for the warning in form 
of a code and an acoustic signal sounds in regular intervals. 
These messages appear every 10 seconds. 
Press the key
to stop the signal
Low level alarm
The circulator is operated without or insufficient bath fluid. 
Switch the unit off with the mains switch, refill bath fluid and switch on! 
Tube breakage has occurred (insufficient filling level of bath fluid caused 
by pumping-out) 
Replace the tubing and refill bath liquid. 
The float is defect (e.g. transport damage). 
Repair by authorized JULABO service personnel.
During the self-test after switch-on a short –circuit is detected between 
pin 2 and pin 4 of the control line or the control line was disconnected 
during operation. 
Reconnect the control line or repair short-circuit.
Excess temperature warning 
or 
Excess temperature alarm 
Type of warning: set to >0 = warning< or >1 = alarm<
Low temperature warning 
or 
Low temperature alarm. 
Typ of warning: set to >0 = warning< or >1 = alarm<
 Cable of working temperature sensor is disconnected or short-
circuited.
Defect of working or excess temperature protector. 
Working temperature sensor and excess temperature protector report a 
temperature difference of more than 35 K. 
Other errors
Internal hardware error – call service