JULABO MA Heating Immersion Circulator User Manual
Page 24
Setting of temperatures
24
Refer to page 32 for switch-over to 3-temperature-mode
CFG = CONFIGURATION
3SP = 3 SETPOINT
Setting of working temperature in the
menu
1. Press the key
. The value >tx<
flashes
2. Select SETPOINT >t 1< or >t 2< or >t 3< using the key
or
.
3. Confirm by pressing the
key.
The circulator uses the new working temperature value for
temperature control.
Example: setting / adjustment of of pre-settings of "t 3"
1. Press
the
key. The parameter >tx< flashes.
2. Select the setpoint >t3< by pressing
or
.
3. Keep the key
pressed until the integer digits flash
.
(example: <70>)
4. Change the value by pressing
and
to 85.00 °C and
confirm by pressing the
key.
The decimal digits flash
and can be adjusted if desired.
Confirm once more by pressing the
key.
Example on the left: SETPOINT >t3< / 85.00.
The end of the adjustment is signaled by the flashing message
>t3<
If the active setpoint (SETPNT) is changed, the new value is
immediately used for the control of the working temperature.
The heater control indicator flashes.
If the other two setpoints (not activated for control) are changed
the
MENU has to be left by pressing the
key after
the decimal digits have been confirmed
Notice: Refer to chapter
9.6. MENU LIMITS