beautypg.com

JULABO MB-5M Heating Circulators with Open Bath User Manual

Page 51

background image

MB

51

14. Cleaning the unit

Caution:

Always turn off the unit and disconnect the mains cable from the power
source before cleaning the unit.

Prevent humidity from entering into the circulator.

Electrical connections and any other work must be performed by qualified
personnel only.


Cleaning:

For cleaning the bath tank and the immersed parts of the circulator, use low
surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface tension
water.

The circulator is designed for continuous operation under normal conditions.
Periodic maintenance is not required.

The tank should be filled only with a bath fluid recommended by JULABO. To
avoid contamination, it is essential to change the bath fluid from time to time.



14.1. Cleaning of bath tanks made of Plexiglas® and Makrolon®

Bath tanks made of Plexiglas® and Makrolon® are used when processes
within the samples have to be visually observed. The transparency of the
bath tanks must not be ruined when cleaning them.

Caution:

Unsuitable cleaning products for bath tanks made of Plexiglas® and
Makrolon®.

These baths tanks are not resistant to solvents and pure alcohol.

Incorrect cleaning products cause opacity of the tank’s surface, cause stress
cracks or dissolve the adhesive. Bath tanks made of Plexiglas® and
Makrolon® will leak.

 Bath tanks have to be wet-cleaned – never rub dry! Cloths and sponges

used for cleaning must not be dirty (abrasive particles, dust).

 Use a dampened microfiber cloth for a good, streak-free cleaning.
 For cleaning use only warm water with a few drops of detergent and a soft

cloth.

 In case of heavy, especially greasy soiling benzene-free clean petrol

(benzine, light petrol) may be used.