JULABO MB-5M Heating Circulators with Open Bath User Manual
Page 30

Safety installations, warning functions
30
8.3.
Over and Sub temperature warning function
Over temperature
t High
205 °C
Sub temperature
t Low
-55 °C
If the observance of a working temperature value >t 3< has to be
supervised for a sensitive temperature application, then set over
and sub temperature warning values. In the example below the
setpoint >t 3< 85 °C is surrounded by the values >t High< 87 °C
and >t Low< 83 °C. The electronics immediately register if the
actual temperature breaches one of the set limit values. The
resulting reaction is defined in the menu item >Aty (ALARM-
TYPE)< refer to (page 29).
Factory settings:
t High
205 °C
t Low
-99.99 °C
1. Press the key
.
2. By pressing the
or
key select the menu > tHigh <
or > tLow <.
3. Press the
key. The integer digits flash.
4. Change the values to 87. °C and/or 83. °C by pressing the
and
key and confirm with the
key.
The decimal digits flash and can be adjusted if desired.
Confirm once more by pressing the
key.
See above examples.
The warning functions are only activated if the actual bath
temperature remains within the set limit values for 3 seconds
after switch-on.
Recommendation:
Set the over temperature warning value > t High < 5 °C to 10 °C
above the working temperature setpoint.
Set the sub temperature warning value > t Low < 5 °C to 10 °C
below the working temperature setpoint.