JULABO MB-5M Heating Circulators with Open Bath User Manual
Page 28

Safety installations, warning functions
28
8.1.1. Early
warning
system, low level protection
Warning:
For refill always use the same bath fluid type that is already in the bath.
(patented)
This low level protection is independent of the control
circuit and is divided in two sections.
1. Switch in stage 1 recognizes a defined
fluid level
.
An audible warning (interval tone) sounds and on
the MULTI-DISPLAY (LED) the message "E 40"
appears.
Refill bath fluid!
2. Switch in stage 2 recognizes a low fluid level
.
If stage 2 of the low level protection device
(according to IEC 61010-2-010) is triggered, a
complete shutdown of the heater and circulating
pump is effected.
A continuous alarm tone sounds and a message
>CODE 01< appears on the MULTI-DISPLAY
(LED).
Turn off the unit with the mains switch, refill bath fluid and turn
the unit on again!
3. Float
4. Circulating
pump
5. Heater