JULABO MB-5M Heating Circulators with Open Bath User Manual
Page 21

MB
21
5.4.
Filling / Draining
Notice:
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit
on for a short moment, for example).
Store and dispose the used bath fluid according to the laws for environmental
protection.
Filling
Take care that no liquid enters the interior of the circulator.
Recommended maximum filling level with water as bath liquid:
20 mm below the tank rim
After filling, immerse the samples in the bath or place the lid on
the bath, in case the opening is not to be used.
The circulator provides an early warning system for low level
that may be triggered when changing samples in the bath.
Draining
Press the mains switch to turn the circulator off.
Remove the circulator from the bath tank.
Empty the bath tank.