ERROR MESSAGES
Description
-05 WORKING SENSOR ALARM
Working temperature sensor short-circuited or
interrupted.
-06 SENSOR DIFFERENCE ALARM
Sensor difference alarm.
Working temperature and safety sensors
report a temperature difference of more than
35 K.
-07 I2C-BUS ERROR
Internal error when reading or writing the I2C
bus.
-08 INVALID COMMAND
Invalid command.
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-12 TEMPERATURE MEASUREMENT
ALARM
Error in A/D converter.
-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the
high and low temperature warning limits. But
value is stored.
-14 EXCESS TEMPERATURE
PROTECTOR ALARM
Excess temperature protection alarm
-30 CONFIGURATION ERROR:
CONFIRM BY PRESSING
CIRCULATOR
The configuration of the circulator does not
conform to its present use.
Press
to automatically perform a single
modification of the configuration.
-33 SAFETY SENSOR ALARM
Excess temperature sensor short-circuited or
interrupted.
-40 NIVEAU LEVEL WARNUNG
Low liquid level warning in the internal
reservoir.