4. Technical specifications
4.1.
F250
Recirculating Cooler
F250
Working temperature range
°C
-10 ... +40
Temperature stability
°C
±0.5
Temperature selection:
digital
via key pad
indication on LED-DISPLAY
Temperature indication:
LED-DISPLAY
Adjustment and display resolution
°C
0.1
Temperature control
PID 1
Temperature sensor
Pt 100
Excess temperature protection
85 °C - fixed value
Low liquid level protection
float switch
Circulating pump:
discharge, max.at 0 bar
l/min
15
pressure, max. at 0 Liters
bar
0.35
Filling level indicator
sight glass
Filling volume
from ... to
Liters
1.7 ... 2.6
Dimensions (WxLxH)
cm
24x40x52
Weight
kg
27.0
Ambient temperature range
°C
5 ... 40
Return flow temperature
max.
°C
80
Cooling compressor
1- stage / air cooled
Refrigerant
R134a
Cooling capacity
at 100 V / 60 Hz
at 115 V / 60 Hz
at 200 V / 60 Hz
at 230 V / 60 Hz
at 230 V / 50 Hz
at 100 V / 50 Hz
at 200 V / 50 Hz
Medium: Mixture water-glycol
°C
W
°C
W
+20 +15 +10 +5
0
-5
-10
250 240 220 210 180
90
60
+20 +15 +10 +5
0
-5
-10
220 210 195 185 170
80
40
Mains power connection
V/ Hz
230
±10 % / 50
Current draw
(at 230 V)
A
3.0
Mains power connection
V/ Hz
230
±10 % / 60
Current draw
(at 230 V)
A
2.0
Mains power connection
V/ Hz
200 -5 %; +21 % / 50-60
Current draw
(at 200 V / 50 Hz)
(at 200 V / 60 Hz)
A
A
2.0
2.1
Mains power connection
V/ Hz
115
±10 % / 60
Current draw
(at 115 V)
A
4.0
Mains power connection
V/ Hz
100
±10 % / 50-60
Current draw
(at 100 V / 50 Hz)
(at 100 V / 60 Hz)
A
A
5.0
4.7
All measurements have been carried out at: rated voltage and frequency