1.951.4806-V1

15

6. Moving up and connect

Risk of injury for hands. Close cover carefully.

Carry the unit with 2 persons.

Wear safety shoes.

6.1.

Transportation and site conditions

F250

Caster platform
(Order No.: 8910045


F500, F1000


Transport

Lifting and Transport:
At F250: Lift the unit with two persons taking hold of

its bottom plate. For transport set the unit on
a suitable caster platform
(Order No.: 8910045).

At F500, F1000: The unit is lifted by two people by the

hand grips (front and back). For transport by
one person, the device can be lifted and
moved forward on the rear casters.

• Place the unit on an even surface on a base made

of nonflammable material.

• Cooling machine, pump motor and electronics

produce intrinsic heat that is dissipated via the
venting openings! Never cover these openings!

• The air vents of the unit must not be covered.
• Keep at least 20 cm of open space at the front and

rear venting grid.

• Do not install the unit in the immediate vicinity of

heat sources and do not expose it to sunlight.

• Ensure good ventilation of the site.
• The place of installation should be large enough and

provide sufficient air ventilation to ensure the room
does not warm up excessively because of the heat
the instrument rejects to the environment. (Max.
permissible ambient temperature: 40 °C).
For a fault (leakage) in the refrigeration system, the
standard EN 378 prescribes a certain room space to
be available for each kg of refrigerant.
The refrigerant quantity is specified on the type
plate.

• For 0.25 kg of refrigerant R134a, 1 m

3

of space is

required.