Moving up and connect, Transportation and site conditions – JULABO F1000 Recirculating Cooler User Manual
Page 15

F250, F500, F1000
1.951.4806-V1
15
6. Moving up and connect
Risk of injury for hands. Close cover carefully.
Carry the unit with 2 persons.
Wear safety shoes.
6.1.
Transportation and site conditions
F250
Caster platform
(Order No.: 8910045
F500, F1000
Transport
• Lifting and Transport:
At F250: Lift the unit with two persons taking hold of
its bottom plate. For transport set the unit on
a suitable caster platform
(Order No.: 8910045).
At F500, F1000: The unit is lifted by two people by the
hand grips (front and back). For transport by
one person, the device can be lifted and
moved forward on the rear casters.
• Place the unit on an even surface on a base made
of nonflammable material.
• Cooling machine, pump motor and electronics
produce intrinsic heat that is dissipated via the
venting openings! Never cover these openings!
• The air vents of the unit must not be covered.
• Keep at least 20 cm of open space at the front and
rear venting grid.
• Do not install the unit in the immediate vicinity of
heat sources and do not expose it to sunlight.
• Ensure good ventilation of the site.
• The place of installation should be large enough and
provide sufficient air ventilation to ensure the room
does not warm up excessively because of the heat
the instrument rejects to the environment. (Max.
permissible ambient temperature: 40 °C).
For a fault (leakage) in the refrigeration system, the
standard EN 378 prescribes a certain room space to
be available for each kg of refrigerant.
The refrigerant quantity is specified on the type
plate.
• For 0.25 kg of refrigerant R134a, 1 m
3
of space is
required.