Autostart on / off – JULABO F1000 Recirculating Cooler User Manual
Page 26

Operating procedures
26
1.951.4806-V1
8.7.
AUTOSTART ON / OFF
 
 
 
 
The recirculating cooler has been configured and supplied by 
JULABO according to N.A.M.U.R. recommendations. This means 
for the start mode, that the unit must enter a safe operating state 
after a power failure (non-automatic start mode). This safe 
operating state is indicated by „OFF“ on the LED temperature 
display. A complete shutdown of the main functional elements 
such as compressor and circulating pump is effected 
simultaneously. 
 
The set values are stored in the memory, and by pressing the main 
switch in manual mode, the device is restored to operation.
 
Should such a safety standard not be required, the AUTOSTART 
function (automatic start mode) may be activated, thus allowing the 
start of the circulator directly by pressing the mains power switch 
or using a timer. 
Keep depressed enter and turn on the unit with the mains power 
switch. 
 
For a short while the LED display indicates the effective start 
mode: 
= AUTOSTART on.
= AUTOSTART off.
Danger from unattended device start. 
For supervised or unsupervised operation with the AUTOSTART 
function, avoid any hazardous situation to persons or property. 
The circulator does no longer conform to N.A.M.U.R. 
recommendations. 