Power connection, Filling – JULABO F1000 Recirculating Cooler User Manual
Page 23

F250, F500, F1000
1.951.4806-V1
23
8.2.
Power connection
Danger of electric shock!
• Only connect the unit to a power socket with earthing contact
(PE – protective earth)! We disclaim all liability for damage
caused by incorrect line voltages!
• The power supply plug serves as safe disconnecting device
from the line and must be always easily accessible.
• Never operate equipment with damaged mains power cables.
• Regularly check the mains power cables for material defects
(e.g. for cracks).
• No liability for incorrect line voltages!
Make sure that the line voltage and frequency match the supply
voltage specified on the type plate.
Max. permissible mains fluctuations, see Technical specifications.
8.3.
Filling
Take care that no liquid enters the interior of the circulating cooler.
Positions as shown in chapter 7.Operating controls and functional
elements, page 18.
Connect the tubing from the external system to the pump
connectors and check for leaks
Check to make sure that the drain screw (Pos. 10 at F250, or
Pos. 12 at F500, F1000) is closed.
•
Remove the cover of filling (Pos. 5).
•
Fill bath fluid to the upper level of the fill level indicator
(Pos. 4).
•
Switch on the unit with the main switch (Pos. 1)
•
Start the unit. Therefore press the key
for about 4
seconds.
•
The bath fluid will be pumped into the externally connected
system. Refill the bath fluid to the upper marking of the level
indicator.
•
The chiller is ready for operation.