Technical specifications

10

1.951.4806-V1

4.4.

Warning functions and safety installations

Excess temperature protection

85 °C - fixed value

Low liquid level protection

float switch

Alarm messages

optical + audible (permanent)

Overload protection

for compressor and pump motor

Classification according to DIN 12876-1

Class I


Environmental conditions according to IEC 61 010-1:
- Use only indoor.
- Altitude up to 2000 m - normal zero.
- Ambient temperature: +5 ... +40 °C
- Air humidity:
- Max. rel. humidity 80 % for temperatures up to +31 °C,
- linear decrease down to 50 % relative humidity at a temperature of +40 °C,
- max. permissible mains fluctuations, see Technical specifications.

The unit corresponds to Class

I

Overvoltage category

II

Pollution degree

2

Fire hazard!
The unit is not for use in explosive atmosphere.

Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).