F500 – JULABO F1000 Recirculating Cooler User Manual
Page 8

Technical specifications
8
1.951.4806-V1
4.2.
F500
Recirculating Cooler
F500
Working temperature range
°C
0 ... +40
Temperature stability
°C
±0.5
Temperature selection:
digital
via key pad
indication on LED-DISPLAY
Temperature indication:
LED-DISPLAY
Adjustment and display resolution
°C
0.1
Temperature control
PID 1
Temperature sensor
Pt 100
Excess temperature protection
85 °C - fixed value
Low liquid level protection
float switch
Circulating pump:
discharge, max.at 0 bar
l/min
24
pressure, max. at 0 Liters
bar
0.5
Filling level indicator
sight glass
Filling volume
from ... to
Liters 5,0 ... 7,5
Dimensions (WxLxH)
cm
37,5x44x59
Weight
kg
34.0
Ambient temperature range
°C
5 ... 40
Return flow temperature
max.
°C
80
Cooling compressor
1- stage / air cooled
Refrigerant
R134a
Cooling capacity
at 230 V / 50 Hz
Medium: Mixture water-glycol
°C
W
+20 +10 +5
0
500 400 300 250
Mains power connection
V/ Hz 230 -10 %; +7 % / 50
Current draw
(at 230 V)
A
2.6
Mains power connection
V/ Hz 230
±10 % / 60
Current draw
(at 230 V)
A
2.5
Mains power connection
V/ Hz 115
±10 % / 60
Current draw
(at 115 V)
A
5.8
Mains power connection
V/ Hz 100 -5 %; +10 % / 50-60
Current draw
(at 100 V / 50 Hz)
(at 100 V / 60 Hz)
A
A
5.9
5.5
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.