Operating procedures, Bath fluids – JULABO F1000 Recirculating Cooler User Manual
Page 22

Operating procedures
22
1.951.4806-V1
8. Operating procedures
8.1. Bath fluids
No liability for use of other bath liquids!
Please contact JULABO before using other than recommended
bath fluids. JULABO takes no responsibility for damages
caused by the selection of an unsuitable bath fluid
Do not use alcohols.
Water:
The quality of water depends on local conditions.
• Due to the high concentration of lime, hard water is not
suitable for temperature control because it leads to
calcification in the bath.
• Ferrous water can cause corrosion - even on stainless
steel.
• Chloric water can cause pitting corrosion.
• Distilled and deionized water is unsuitable. Their special
properties cause corrosion in the bath, even in stainless
steel.
• No liablity for use with water. Danger of freezing at working
temperatures <5 °C.
Mixture water -glycol:
Strictly observe the safety data and handling instructions from
the manufacturer.
The proportion of water might evaporate by and by. Check the
mixing ratio regularly and refill water if necessary.
Recommended bath fluids:
Bath fluid
Temperature range
JULABO Thermal G
-30 °C ... 80 °C
Water/Glycole (50:50)
-30 °C ... 50 °C
soft/decalcified water
+5 °C ... 80 °C
Order No.
Ordering text
Volume
8 940 124
8 940 125
JULABO Thermal G
JULABO Thermal G
10 Liter
5 Liter
See website for list of recommended bath fluids.
Contact: www.julabo.com
Use of non-recommended bath fluids may result in a fire
hazard or other hazard:
JULABO will assume no liability for damages resulting from use of