beautypg.com

F1000 – JULABO F1000 Recirculating Cooler User Manual

Page 9

background image

F250, F500, F1000

1.951.4806-V1

9

4.3.

F1000

Recirculating Cooler

F1000

Working temperature range

°C

0 ... +40

Temperature stability

°C

±0.5

Temperature selection:

digital

via key pad

indication on LED-DISPLAY

Temperature indication:

LED-DISPLAY

Adjustment and display resolution

°C

0.1

Temperature control

PID 1

Temperature sensor

Pt 100

Excess temperature protection

85 °C - fixed value

Low liquid level protection

float switch

Circulating pump:

discharge, max.at 0 bar

l/min

23

pressure, max. at 0 Liters

bar

1.0

Filling level indicator

sight glass

Filling volume

from ... to

Liter

7.0 ... 9.5

Dimensions (WxLxH)

cm

37.5x49x64

Weight

kg

46

Ambient temperature range

°C

5 ... 40

Return flow temperature

max.

°C

80

Cooling compressor

1- stage / air cooled

Refrigerant

R134a

Cooling capacity

at 230 V / 50 Hz

Medium: Water-glycol


°C
W


+20 +10

+5

0

1000 700

550

350

Mains power connection

V/ Hz

230

±10 % / 50

Current draw

(at 230 V / 50 Hz)

A

5

Mains power connection

V/ Hz

230

±10 % / 60

Current draw

(at 230 V / 60 Hz)

A

4

Mains power connection

V/ Hz

200

±10 % / 50-60

Current draw

(at 200 V / 50 Hz)

A

-

Current draw

(at 200 V / 60 Hz)

A

-

Mains power connection

V/ Hz

115

±10 % / 60

Current draw

(at 115 V / 60 Hz)

A

9


All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.