F1000 – JULABO F1000 Recirculating Cooler User Manual
Page 9
F250, F500, F1000
1.951.4806-V1
9
4.3.
F1000
Recirculating Cooler
F1000
Working temperature range
°C
0 ... +40
Temperature stability
°C
±0.5
Temperature selection:
digital
via key pad
indication on LED-DISPLAY
Temperature indication:
LED-DISPLAY
Adjustment and display resolution
°C
0.1
Temperature control
PID 1
Temperature sensor
Pt 100
Excess temperature protection
85 °C - fixed value
Low liquid level protection
float switch
Circulating pump:
discharge, max.at 0 bar
l/min
23
pressure, max. at 0 Liters
bar
1.0
Filling level indicator
sight glass
Filling volume
from ... to
Liter
7.0 ... 9.5
Dimensions (WxLxH)
cm
37.5x49x64
Weight
kg
46
Ambient temperature range
°C
5 ... 40
Return flow temperature
max.
°C
80
Cooling compressor
1- stage / air cooled
Refrigerant
R134a
Cooling capacity
at 230 V / 50 Hz
Medium: Water-glycol
°C
W
+20 +10
+5
0
1000 700
550
350
Mains power connection
V/ Hz
230
±10 % / 50
Current draw
(at 230 V / 50 Hz)
A
5
Mains power connection
V/ Hz
230
±10 % / 60
Current draw
(at 230 V / 60 Hz)
A
4
Mains power connection
V/ Hz
200
±10 % / 50-60
Current draw
(at 200 V / 50 Hz)
A
-
Current draw
(at 200 V / 60 Hz)
A
-
Mains power connection
V/ Hz
115
±10 % / 60
Current draw
(at 115 V / 60 Hz)
A
9
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.