3 I/O Assembly Object
Class Code 04
h
A collection of assembly objects allows the sending of attributes from different
application objects in one message (i.e. Polling I/O).
The following table identifies the I/O assembly instances supported by the gauge
device.
Number
Type
Name
1
Input
Pressure Value (Active Instance)
2
Input
Exception Status and INT Pressure Value (Active In-
stance)
4
Input
REAL Pressure Value (Active Instance)
5
Input
Exception Status and REAL Pressure Value (Active
Instance)
8
Input
Exception Status
9
Input
Active Instance, Active Pressure Value
10
Input
Exception Status and Active Instance and INT Active
Pressure Value
12
Input
Active Instance
REAL Active Pressure Value
13
Input
Exception Status
Active Instance
REAL Active Pressure Value