To perform the Full Scale Adjust Request vent the gauge to atmosphere and then
start the Full Scale Adjust Service with a target value as defined below .

BPG400-SD:

Para-
meter

Data
Type

The target value for the full scale calibration

Data
Unit

Counts

mbar *)

Pascal *)

Torr *)

Target
Value

INT

0x7918

0x03E8

0x7FFF

0x02ED

REAL 0x46F23000

0x447A0000

0x47C35000

0x443B83F3

HPG400-SD:

Para-
meter

Data
Type

The target value for the full scale calibration

Data
Unit

Counts

mbar *)

Pascal *)

Torr *)

Target
Value

INT

0x767D

0x03E8

0x7FFF

0x02ED

REAL 0x46ECFA00

0x447A0000

0x47C35000

0x443B83F3

*)

INFICON recommends to use mbar, Torr or Pascal only with the data

type REAL.

Special limitations apply for INT:
The full scale value of the gauge is 1000 mbar

10000 Pascal. But the

max. pressure of the gauge can not be described using INT and Pascal.
The target value 0x7FFF (the maximum INT value) allows to perform a Full
Scale Adjust Service for the unit Pascal.
For the data type INT and pressure values below 1 (in the actual pressure
unit) you always get a pressure of zero.

Service

Code

Service Name

Description of Service

61

h

Set Degas State Activates/deactivates degas mode according to

the parameter Degas State. Degas mode may
be terminated either automatically by device
timeout (3 min) or remotely by this service.

Parameter

Data Type

Description

Semantics of Values

Degas
State

BOOL

0

switches Degas OFF

1

switches Degas ON

The following behavior with respect to Data Type applies:
The Data Type value will be set automatically based upon the first valid I/O con-
nection established by the device.
If no established I/O connections exist, which include an attribute from this object,
then the Data Type attribute is settable provided that the object is in the Idle State.

The following example demonstrates this behavior:
A device specifies an instance of the S-Analog Sensor object as well as two static
Assembly object instances, both with data attribute components mapped to this
object instance. Assembly object instance ID 1 specifies INT data types and As-
sembly object instance ID 2 specifies REAL data types.

Full Scale adjust algorithm

Full Scale Adjust Request
Service Data Field Parame-
ters

2.4.5 Object-Specific Services

on Instance 2 / Hot
Cathode Ion Gauge
(Only BPG400-SD)

Set Degas State Request
Service Data Field
parameters

2.4.6 Behavior

Data Type