INFICON HPG400-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual
Page 25

tira03e1-b (0310) BPG/HPG400-SDv1.cp
25
To perform the Full Scale Adjust Request vent the gauge to atmosphere and then
start the Full Scale Adjust Service with a target value as defined below .
BPG400-SD:
Para-
meter
Data
Type
The target value for the full scale calibration
Data
Unit
Counts
mbar *)
Pascal *)
Torr *)
Target
Value
INT
0x7918
0x03E8
0x7FFF
0x02ED
REAL 0x46F23000
0x447A0000
0x47C35000
0x443B83F3
HPG400-SD:
Para-
meter
Data
Type
The target value for the full scale calibration
Data
Unit
Counts
mbar *)
Pascal *)
Torr *)
Target
Value
INT
0x767D
0x03E8
0x7FFF
0x02ED
REAL 0x46ECFA00
0x447A0000
0x47C35000
0x443B83F3
*)
INFICON recommends to use mbar, Torr or Pascal only with the data
type REAL.
Special limitations apply for INT:
The full scale value of the gauge is 1000 mbar
10000 Pascal. But the
max. pressure of the gauge can not be described using INT and Pascal.
The target value 0x7FFF (the maximum INT value) allows to perform a Full
Scale Adjust Service for the unit Pascal.
For the data type INT and pressure values below 1 (in the actual pressure
unit) you always get a pressure of zero.
Service
Code
Service Name
Description of Service
61
h
Set Degas State Activates/deactivates degas mode according to
the parameter Degas State. Degas mode may
be terminated either automatically by device
timeout (3 min) or remotely by this service.
Parameter
Data Type
Description
Semantics of Values
Degas
State
BOOL
0
switches Degas OFF
1
switches Degas ON
The following behavior with respect to Data Type applies:
The Data Type value will be set automatically based upon the first valid I/O con-
nection established by the device.
If no established I/O connections exist, which include an attribute from this object,
then the Data Type attribute is settable provided that the object is in the Idle State.
The following example demonstrates this behavior:
A device specifies an instance of the S-Analog Sensor object as well as two static
Assembly object instances, both with data attribute components mapped to this
object instance. Assembly object instance ID 1 specifies INT data types and As-
sembly object instance ID 2 specifies REAL data types.
Full Scale adjust algorithm
Full Scale Adjust Request
Service Data Field Parame-
ters
2.4.5 Object-Specific Services
on Instance 2 / Hot
Cathode Ion Gauge
(Only BPG400-SD)
Set Degas State Request
Service Data Field
parameters
2.4.6 Behavior
Data Type