INFICON HPG400-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual
Page 10

10
tira03e1-b (0310) BPG/HPG400-SDv1.cp
Attribute ID
Access Rule NV/
V
Name
DeviceNet Data Type
Description of Attribute
3
Get
NV DeviceType
SHORT STRING
ASCII Text, "CG" combination gauge
4
Get
NV SEMI Standard
Revision Level
SHORT STRING
Specifies the revision level of the SEMI
S/A network standard to which the device
complies (for example "E54-0997").
5
Get
NV Manufacturer’s Name
SHORT STRING
"INFICON"
6
Get
NV Manufacturer’s Model
Number
SHORT STRING
for example: BPG400-SD
7
Get
NV Software Revision
Level
SHORT STRING
ASCII Text, for example "xxxyyy"
xxx
version of the measuring print
yyy
version of the DeviceNet print
8
Get
NV Hardware Revision
Level
SHORT STRING
ASCII text, for example "1.001"
9
Get
NV Manufacturer’s Serial
Number
SHORT STRING
ASCII text
10
Get
NV Device Configuration
SHORT STRING
Depending on device configuration
(
→ Appendix B)
11
Get
V
Device Status
USINT
→ "Semantics" section below.
12
Get
V
Exception Status
BYTE
→ "Semantics" section below.
13
Get
V
Exception Detail
Alarm
STRUCT of:
A structure of three structures containing
a bit mapped representation of the alarm
detail
Common
Exception Detail
STRUCT of:
Size 2
USINT
Number of common detail bytes
Detail 0
ARRAY of:
→ "Semantics" section below.
Detail 1
BYTE
→ "Semantics" section below.
Device Exception
Detail
STRUCT of:
Size 4
USINT
Number of device detail bytes
Detail 0
ARRAY of:
S-Analog Sensor Instance 1 (Pirani)
Sensor alarm byte 0
Detail 1
BYTE
S-Analog Sensor Instance 1 (Pirani)
Sensor alarm byte 0
Detail 2
S-Analog Sensor Instance 2
(Hot cathode) Sensor Alarm byte 0
Detail 3
S-Analog Sensor Instance 2
(Hot cathode) Sensor alarm byte 1
Manufacturer
Exception Detail
STRUCT of:
Size 1
USINT
Number of Manufacturer detail bytes
Detail 0
ARRAY of:
Serial communication alarm
2.3.2 Instance Attributes