Attribute ID

Access Rule NV/

V

Name

DeviceNet Data Type

Description of Attribute

14

Get

V

Exception Detail
Warning

STRUCT of:

A structure of three structures containing
a bit mapped representation of the
warning detail

Common Exception
Detail

STRUCT of:

Size 2

USINT

Number of Common Detail Bytes

Detail 0

ARRAY of:

→ "Semantics" section below.

Detail 1

BYTE

→ "Semantics" section below.

Device Exception
Detail

STRUCT of:

Size 6

USINT

Number of Device Detail Bytes

Detail 0

ARRAY of:

S-Analog Sensor Instance 1 (Pirani)
Status Extension

Detail 1

BYTE

S-Analog Sensor Instance 1 (Pirani)
Sensor Warning Byte 0

Detail 2

BYTE

S-Analog Sensor Instance 1 (Pirani)
Sensor Warning Byte 1

Detail 3

S-Analog Sensor Instance 2
(hot cathode) Status Extension

Detail 4

S-Analog Sensor Instance 2
(hot cathode) Sensor Warning Byte 0

Detail 5

S-Analog Sensor Instance 2
(hot cathode) Sensor Warning Byte 1

Manufacturer
Exception Detail

STRUCT of:

Size 1

USINT

Number of Manufacturer Detail Bytes

Detail 0

Byte

serial common warning

15

Set

NV Alarm Enable

BOOL

→ "Semantics" section below.

16

Set

NV Warning Enable

BOOL

→ "Semantics" section below.

101

Get

NV Sensitivity

USINT

→ "Semantics" section below.
only HPG400-SD

102

Get

NV Emission On

Threshold

INT or REAL

→ "Semantics" section below.
only HPG400-SD

This attribute represents the current state of the device. Its value changes as the
state of the device changes. The following values are defined:

Attribute Value

State

0

Undefined

1

Self Testing

2

Idle

3

Self-Test Exception

4

Executing

5

Abort

6

Critical Fault

In case of a Self-Test Exception restart the device with a reset out of the box
(Identity Object Class 1, Instance 1, Service 5, Service target value: 1).

2.3.2.1 Semantics