beautypg.com

INFICON Stripe CDG045Dhs 0.01 ... 1000Torr / mbar, new 10mTorr User Manual

Page 30

background image

30

tina84e1 (2014-07)

4.3.1

Adjusting the Setpoints SP1, SP2

The thresholds of the setpoints can be adjusted via

• the buttons on the gauge
• the diagnostic port (→  [4])
• the EtherCAT interface (→  [3]).

DANGER

DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing an button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the button only if you are sure that no
malfunction will cause.

The setpoints can be adjusted via

• the buttons on the gauge,
• the diagnostic port (→  [4])
• the EtherCAT interface (→  [3]).

Adjusting Setpoint <1>

Push the button with a pin (max. ø1.1 mm). The

gauge changes to the switching function mode and outputs
the current threshold value at the measurement value
output for about 10 s (LED <1> blinks).