Software update 6.6.1, Updating the software of the control unit, Updating the software of the ms module – INFICON CU1000 Control unit User Manual
Page 22: Updating the software of the i/o module, 6 software update, 1 updating the software of the control unit, 2 updating the software of the ms module, 3 updating the software of the i/o module
22
Operation
6.6
Software update
6.6.1
Updating the software of the control unit
1
Copy these files to the main directory of a USB stick:
– Handset_IFC_Vx.x.xx.exe
– Handset_IFC_Vx.x.xx.key
2
Plug the USB stick into the USB port on the control unit.
3
Select on measurement screen:
F
UNCTIONS
> D
ATA
> U
PDATE
> C
ONTROL
UNIT
4
Press O
K
to confirm message.
5
Press S
TART
to begin.
6.6.2
Updating the software of the MS module
1
Copy these files to the main directory of a USB stick:
– Flash_LDS3000_MSB_Vxx.xx.xxx.bin
2 Plug the USB stick into the USB port on the control unit.
3
Select on measurement screen:
F
UNCTIONS
> D
ATA
> U
PDATE
> MSB
4
Press O
K
to confirm message.
5
Press S
TART
to begin.
6.6.3
Updating the software of the I/O module
Information The software of the I/O module can be updated from the control unit if
the mass spectrometer module has the software version “MS-Modul
1.02” or higher.
1
Copy these files to the main directory of a USB stick:
– Flash_LDS3000_IO_Vxx.xx.xxx.bin
2
Switch DIP switch S2-3 of the I/O module from OFF to ON (rising flank).
3
Plug the USB stick into the USB port on the control unit.
4
Select on measurement screen:
F
UNCTIONS
> D
ATA
> U
PDATE
> I/O
MODULE
5
Press O
K
to confirm message.
6
Press S
TART
to begin.
7
Switch DIP switch S2-3 of the I/O module to OFF.