22

Operation

6.6

Software update

6.6.1

Updating the software of the control unit

1

Copy these files to the main directory of a USB stick:

– Handset_IFC_Vx.x.xx.exe

– Handset_IFC_Vx.x.xx.key

2

Plug the USB stick into the USB port on the control unit.

3

Select on measurement screen:
F

UNCTIONS

> D

ATA

> U

PDATE

> C

ONTROL

UNIT

4

Press O

K

to confirm message.

5

Press S

TART

to begin.

6.6.2

Updating the software of the MS module

1

Copy these files to the main directory of a USB stick:
– Flash_LDS3000_MSB_Vxx.xx.xxx.bin

2 Plug the USB stick into the USB port on the control unit.

3

Select on measurement screen:
F

UNCTIONS

> D

ATA

> U

PDATE

> MSB

4

Press O

K

to confirm message.

5

Press S

TART

to begin.

6.6.3

Updating the software of the I/O module

Information The software of the I/O module can be updated from the control unit if

the mass spectrometer module has the software version “MS-Modul
1.02” or higher.

1

Copy these files to the main directory of a USB stick:

– Flash_LDS3000_IO_Vxx.xx.xxx.bin

2

Switch DIP switch S2-3 of the I/O module from OFF to ON (rising flank).

3

Plug the USB stick into the USB port on the control unit.

4

Select on measurement screen:
F

UNCTIONS

> D

ATA

> U

PDATE

> I/O

MODULE

5

Press O

K

to confirm message.

6

Press S

TART

to begin.

7

Switch DIP switch S2-3 of the I/O module to OFF.