beautypg.com

Ambient conditions, Factory settings – INFICON CU1000 Control unit User Manual

Page 11

background image

Description

11

4.3.3

Ambient conditions

4.3.4

Factory settings

The parameters of the control unit are stored in the control unit.

You can use the control unit to rest the parameters of the mass spectrometer module
and the parameters of the control unit to the factory settings, independent of each
other (see

Chapter 6.2.3, page 19

).

The factory settings of the mass spectrometer module are listed in the installation
manual of the mass spectrometer module.

Permissible ambient temperature
(during operation)

10 °C ... 45 °C

Permissible storage temperature

-20 °C ... 60 °C

Max. relative
humidity

< +31 °C

80%

+31 °C to +40 °C

decreasing linearly from 80% ... 50%

> +40 °C

50%

Type of protection

IP 20

Pollution degree

II

Max. altitude above sea level

2000 m

Parameter

Factory setting

Displ. upper limit exponent

-1

Displ. lower limit exponent

-12

Display off after

1h (= 1 hour)

Display brightness

100%

Measurement view mode

Diagram

Audio alarm mode

Proportional

Automatic scaling

On

Data recording

Off

Decades

3

Pressure unit

mbar

Favorite 1

Volume level

Favorite 2

Measurement view mode

Favorite 3

CAL

Error info Operator

No. and text

Error info Supervisor

No., text and info

Error info Viewer

Number only

Used trigger

1

Volume level

8

Leak rate unit SNIFF

mbar*l/s

Leak rate unit VAC

mbar*l/s