Page 21: Lock all access levels, 3 loading or saving parameters, 4 copying or deleting measured data, 5 lock all access levels, 5 info
Operation
21
Information A key switch can be used instead of the PIN. The key switch is
connected to the I/O module via three digital inputs.
6.2.5
Info
I
NFO
> M
EASUREMENT
/T
EMPERATURE
/E
NERGY
/H
ISTORY
/C
ONTROL
UNIT
/MS
MODULE
/
I
NTERFACES
6.3
Loading or saving parameters
You can use a USB stick to backup and restore the control unit and mass
spectrometer module parameters.
► Load parameters as needed:
F
UNCTIONS
> D
ATA
> P
ARAMETER
> L
OAD
> L
OAD
PARAMETER
► Save parameters as needed:
F
UNCTIONS
> D
ATA
> P
ARAMETER
> S
AVE
> S
AVE
PARAMETER
6.4
Copying or deleting measured data
► Copy measured data from the control unit’s internal memory to a USB stick as
needed:
F
UNCTIONS
> D
ATA
> R
ECORDER
> C
OPY
> C
OPY
FILES
► Delete measured data from the control unit’s internal memory as needed:
F
UNCTIONS
> D
ATA
> R
ECORDER
> D
ELETE
> D
ELETE
FILES
6.5
Lock all access levels
► Lock all access levels as needed:
A
CCESS
C
TRL
> V
IEWER
Access Ctrl
Login with PIN
Parameter
Level
Designation
query
set
1
Viewer (V
IE
)
No
Yes
No
2
Operator (O
PE
)
Yes
Yes
Yes
3
Supervisor (S
UP
)
Yes
Yes
Yes
4
Integrator (I
NT
)
Yes
Yes
Yes
Diverse parameters
Information on diverse parameters, e. g. on the pre-amplifier
or the ion source
-