Operation

19

S

ETTINGS

> S

ET

UP

> C

ONTROL

UNIT

> A

UDIO

S

ETTINGS

> S

ET

UP

> C

ONTROL

UNIT

> E

NERGY

Information If the display shows a warning or an error, then a two-pitch signal

sounds simultaneously.

6.2.3

Functions

F

UNCTIONS

> D

ATA

> P

ARAMETERS

> R

ESET

Volume level

Volume of the headphones or active speaker
0 ... 15

Audio alarm mode

Type of audio alarm
No function

-

Proportional

The frequency of the audible signal is

proportional to the bar graph display or

diagram height. The frequency range is

300 Hz to 3300 Hz.

Setpoint

The pitch is proportional to the leakage rate.

A signal sounds if the leakage rate exceeds

the selected trigger value.

Pinpoint

The sound of the acoustic signal changes its
frequency within a specific range of leakage
rates. Range: A decade below the selected
trigger threshold up to one decade above.
The sound keeps at a constant low and a
constant high frequency below and above
this range, respectively.

Trigger

If the selected trigger threshold is exceeded,
a two-pitch signal sounds.

Display off after

Time after which the display switches off to save energy
30 s, 1 min, 2 min, 5 min, 10 min, 30 min, 1 h, ∞ (= never)

Control unit settings Resetting the control unit parameters to the factory settings

-

MSB settings

Resetting the mass spectrometer module parameters to the
factory settings
-

Parameter access
level

Resetting the parameter access levels to the factory settings
-