Information If the display shows a warning or an error, then a two-pitch signal
sounds simultaneously.
6.2.3
Functions
F
UNCTIONS
> D
ATA
> P
ARAMETERS
> R
ESET
Volume level
Volume of the headphones or active speaker
0 ... 15
Audio alarm mode
Type of audio alarm
No function
-
Proportional
The frequency of the audible signal is
proportional to the bar graph display or
diagram height. The frequency range is
300 Hz to 3300 Hz.
Setpoint
The pitch is proportional to the leakage rate.
A signal sounds if the leakage rate exceeds
the selected trigger value.
Pinpoint
The sound of the acoustic signal changes its
frequency within a specific range of leakage
rates. Range: A decade below the selected
trigger threshold up to one decade above.
The sound keeps at a constant low and a
constant high frequency below and above
this range, respectively.
Trigger
If the selected trigger threshold is exceeded,
a two-pitch signal sounds.
Display off after
Time after which the display switches off to save energy
30 s, 1 min, 2 min, 5 min, 10 min, 30 min, 1 h, ∞ (= never)
Control unit settings Resetting the control unit parameters to the factory settings
-
MSB settings
Resetting the mass spectrometer module parameters to the
factory settings
-
Parameter access
level
Resetting the parameter access levels to the factory settings
-