Access ctrl, 4 access ctrl – INFICON CU1000 Control unit User Manual
Page 20
20
Operation
F
UNCTIONS
> D
ATA
> R
ECORDER
> S
ETTINGS
F
UNCTIONS
> D
ATA
> P
ARAMETERS
6.2.4
Access Ctrl
A
CCESS
C
TRL
> O
PERATOR
/S
UPERVISOR
/I
NTEGRATOR
> E
RROR
INFO
A
CCESS
C
TRL
> O
PERATOR
/S
UPERVISOR
/I
NTEGRATOR
A
CCESS
C
TRL
> O
PERATOR
/S
UPERVISOR
/I
NTEGRATOR
> PIN
ASSIGN
Data recording
Activation or deactivation of data recording
Off, On
Record interval
Time interval between the recording of data sets
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s
Memory location
Memory location of the measured data that is recorded
(USB stick, control unit)
USB
The data are stored on a USB stick as a TXT
file. Each TXT file contains the following
information:
- Creation date
- Software versions
- Serial number
- Start time
- Time stamp (measurement indicates offset
in seconds in relation to start time)
- File name
- Time stamp (offset in seconds in relation to
start time)
- Leakage rate (in used unit)
- Pressure P1 (in used unit)
- Device status
internal
The memory in the control unit is limited
(sufficient for recording data for a maximum
of 24 hours).
Recommendation: Record directly to USB
stick.
List
List of parameters that can be changed (sorted
alphabetically)
-
Error info Viewer/
Operator/Supervisor
Depth of error information
Error number, Short text, Extended information
Para. access
List of parameter access levels
-
Operator/Supervisor/
Integrator
PIN for the access levels
0000 ... 9999