6.5 Replacing the overpressure safety relief device at the condenser
Overpressure safety relief device
...................................................................... 23-30-01-04-98
Round bottom flask 500 ml, coated
...................................................................... ...514-83000-02
➨ Remove joint clip at the catchpot.
➨
Unscrew the four Torx screws at the counter holder of the
condenser and remove condenser. Thereby remove the
adapter from the inlet of the condenser.
➨ Pull off old overpressure safety relief device and install
new one. Check for correct position of the PTFE-foil under
the overpressure safety relief device.
➨ Reassembly in reverse order.
➨ Position adapter in the inlet of the condenser and screw
condenser with counter holder to the pump (Torx screws).
Fix with union nut.
➨ Reassemble catchpot.