5 Troubleshooting
Fault
Possible cause
Remedy
❑ Pump does not start or
stops immediately.
➨ Supply voltage too low or
power supply failure?
✔ Check supply voltage. Check
fuse.
➨ Control signal for motor speed
is missing?
✔ Check control signal.
➨ Pressure in outlet pipeline too
high?
✔ Remove blockade in line,
open valve.
➨ Motor overloaded?
✔ Allow motor to cool down,
identify and eliminate cause
of failure. Manual reset is
necessary. Switch off pump
or unplug mains.
❑ Pump does not achieve
its ultimate vacuum or
usual pumping speed.
➨ Leak in the pipeline or vacu-
um system?
✔ Check pump directly - con-
nect vacuum gauge directly
at pump inlet - then check
connection, pipeline and
vacuum system if necessary.
➨ Long, narrow line?
✔ Use lines with larger di-
ameter, length as short as
possible.
➨ Pump has been exposed to
condensate?
✔ Allow pump to run for some
minutes with atmospheric
pressure at the inlet.
➨ Deposits have been formed
inside the pump?
✔ Clean and inspect the pump
heads.
➨ Diaphragms or valves dam-
aged?
✔ Replace diaphragms and/or
valves.
➨ Outgassing substances or va-
por generated in the process?
✔ Check process parameters.
➨ Pump temperature too high
(motor speed reduced)?
✔ Ensure sufficient cooling of
the pump or reduce inlet pres-
sure.
❑ Pump too noisy.
➨ Loud exhaust noise?
✔ Connect hose or silencer to
pump outlet.
➨ Diaphragm crack or dia-
phragm clamping disc loose?
✔ Perform maintenance.
➨ Other than above mentioned
causes?
✔ Contact local distributor.
❑ Pump seized.
✔ Contact local distributor.