Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 01-005-005-36-1

page 20 of 35

Translation of the original instructions

In case of condensation: Check liquid level in the catchpot during operation. Avoid

overflowing of the catchpot.

Do not allow the catchpot to get overfilled. Maximum liquid level approx. 80% to
avoid problems when removing the catchpot.
Check liquid level in the catchpot regularly and drain catchpot in time.

Permissible range of coolant temperature at the exhaust waste vapor condenser:
-15°C to +20°C.

Check hose connections prior to starting operation of the cooling system.
Check coolant hoses regularly during operation.

Removing the catchpot:
Catchpot at the outlet:
Remove joint clip, remove catchpot and drain condensate.

Catchpot at the inlet:
Admit air or inert gas (via the pump inlet) to restore atmospheric pressure in the
catchpot before attempting removal. Remove joint clip. Remove catchpot and drain
condensate.

Reattach drained catchpots.

+ Important: Comply with regulations when disposing of solvents/condensates.

Recycle if possible; purify if contaminated.

4.6 Shutdown

Short-term:
Has the pump been exposed to condensate?
- Allow the pump to continue to run at atmospheric pressure for a few minutes
Has the pump been exposed to media which may damage the pump materials or
forms deposits?
- Check and clean pump heads if necessary.

Long-term:
- Take measures as described in section short-term shutdown.
- Separate pump from the application.
- Drain catchpot.
- Close inlet and outlet port (e. g. with transport caps).
- Close gas ballast.
- Store the pump in dry conditions.

NOTICE

NOTICE