Discontinue regulation, aerate rotary evaporator.
 
 
7.5. Calibrate pressure sensor 
 
The VAC control automatic pressure sensor may be calibrated. 
 
To this end you need an accurate vacuum gauge which is installed in the vacuum line between 
vacuum pump and VAC control automatic vacuum fitting. 
 
7.5.1. Calibrate atmospheric pressure 
 
Remove vacuum hose from VAC control automatic. 
 
 
Turn VAC control automatic power switch OFF. Depress "Set" key, hold 
depressed while turning power switch ON. The "Act Vac" display shows 
your firmware's version number. 
 
Depress "Set" key, "CAL" is displayed.
 
Depress arrow key "Up", present atmospheric pressure appears in the 
"Act Avp" display; "Set Vac" display shows "HI" 
 
Copy pressure gauge reading with arrow keys
 
 
Complete calibration by hitting the "Set" key. 
 
 
7.5.2. Low value calibration 
 
Connect vacuum hose with VAC control automatic and start vacuum pump. Pressure decreases to 
lowest possible level as achieved by the vacuum pump. 
 
If vacuum pump does not reach a pressure of less than 100 mbar, calibration is 
discontinued; an error report appears in the Act Vac display. 
 
Turn power switch OFF. Depress "Set" key, hold and turn power switch 
ON. The "Act Vac" display shows your firmware's version number. 
 
 
Depress "Set" key, "CAL" is displayed. 
 
 
Depress arrow key "down", present pressure appears in the "Act Vac" 
display. "Set Vac" display shows "LO". 
 