Troubleshooting Chart
Problem
Possible Causes
Corrective Action
Unit does not run 
(Digital Display is 
blank) 
No power to unit 
 
Refrigerated units only 
Electrical cord between 
Control Head and 
Refrigeration Power Module 
loose or unplugged 
Power Switch / Circuit 
Breaker in OFF position 
Check that the electrical cord is secure and 
connected to an operating electrical outlet. 
 
Check that the electrical cord is secure and 
properly connected. 
 
 
Place Power Switch / Circuit Breaker in ON 
position. 
Unit does not run 
(STANDBY appears 
on Digital Display) 
Unit in Standby mode
Press Power Key on front panel.
No fluid circulation
Insufficient fluid in reservoir 
Pump impeller jammed 
Add fluid to reservoir. 
Inspect pump and remove debris as required. 
Insufficient circulation Fluid viscosity too high
External tubing diameter 
too small 
Low line voltage 
Replace with lower viscosity bath fluid. 
Replace with larger diameter tubing. 
 
Check and correct as required. 
Unit does not heat
Insufficient fluid in reservoir 
Temperature set point too 
low 
Safety Set Temperature too 
low 
Add fluid to reservoir. 
Increase temperature set point. 
 
Increase Safety Set temperature. 
Insufficient heating
Insufficient circulation 
Low line voltage 
Ambient temperature too 
cool 
Excessive heat loss 
See Insufficient circulation, above. 
Check and correct as required. 
Increase ambient temperature or relocate unit. 
 
Check for heat loss from external tanks and 
hoses; 
Check for vapor/heat loss from internal 
reservoir. 
Temperature 
unstable 
Insufficient circulation 
Debris or mineral build-up 
on pump, heater, or 
temperature sensor. 
Check pump flow and operation. 
Clean as required. 
Unit does not cool
Dust build up on air filter or 
condenser 
Blocked air ventilation 
screens 
Temperature set point is 
too high 
Excessive heat load 
 
Ambient air temperature too 
high (>35°C / 95°F) 
Low or high line voltage 
Clean air filter and/or condenser as required. 
 
Remove blockages as required. 
 
Decrease temperature set point. 
 
Check that heat load does not exceed capacity 
of bath; correct as required. 
Decrease ambient air temperature. 
 
Check and correct as required. 