Setting the Offset Calibration

This sub-menu allows you to match the Circulator’s temperature display to an external reference
thermometer. A value from -3.0° to +3.0°C may be entered; the factory default is 0.0°C.

IMPORTANT:

To prevent the Offset Calibration value from being changed unintentionally, the

following power down/power up sequence is required to enable the Offset Calibration function.

Refrigerating Circulators:

1. Place the power switch/circuit breaker on the rear of the unit in the OFF position.

2.

Return the power switch/circuit breaker to the ON position while pressing and holding the

key.

3.

When STANDBY appears on the display, release the

key and press

.

You can now

proceed as outlined below.

To Access:

Press the key until OFFSET CALIBRATE appears on the display.

To Change:

Press or .

To Accept:

Press or allow the LCD to return to the main operational display

(approximately 10 seconds).

CAUTION:

The Offset Calibration value is always shown in degrees C, even if degrees F is selected as

the temperature unit in which the control and actual bath temperatures are displayed. Your Circulator
will automatically convert the °C offset calibration value to the correct °F display offset value.

Appears when in

Offset Calibrate sub-menu