Setting the Low Limit Temperature
This sub-menu allows you to limit how low the temperature set point may be set. It also serves as a low
limit safety, alerting you if bath temperature falls below the low limit temperature setting. The Low Limit
value may be set from -55° to +10°C / -65° to +50°F; the factory default is -50°C / -60°F.
To avoid unwanted alarms or shutdowns during regular operation, the Low Limit value should be set at
least 5° lower than the selected control temperature.
To Access:
Press the key until LOW LIMIT appears on the display.
To Change:
Press or .
To Accept:
Press or allow the LCD to return to the main operational display
(approximately 10 seconds).
Appears when in the Low
Limit sub-menu
See also other documents in the category Heidolph Equipment:
- Hei-VAP Precision (59 pages)
- Hei-VAP Value (84 pages)
- Rotacool Mini (104 pages)
- Rotacool (15 pages)
- ROTAVAC valve control (188 pages)
- ROTAVAC valve control (174 pages)
- ROTAVAC vario control (34 pages)
- ROTAVAC Vario pumping unit (61 pages)
- LABOROTA 4000 eco (168 pages)
- LABOROTA 4000 up to 4003 (364 pages)
- Heating bath HB 4000 (52 pages)
- LABOROTA 20 eco (180 pages)
- LABOROTA 20 compact and safety (284 pages)
- LABOROTA 20 automatic (190 pages)
- ROTAVAC 20 (35 pages)
- Automatic Module Distimatic (52 pages)
- Synthesis 1 (268 pages)
- MR Hei-Standard (39 pages)
- MR Hei-End (124 pages)
- MR Hei-Mix L (24 pages)
- MR Hei-Mix S (34 pages)
- EKT Hei-Con (78 pages)
- Heat-On (13 pages)
- StarFish Multi-Experiment Work Station (17 pages)
- RZR 1 (44 pages)
- RZR 2020 (58 pages)
- RZR 2051 control (68 pages)
- REAX top (44 pages)
- REAX 2 (48 pages)
- REAX 20 (54 pages)
- Multi Reax (60 pages)
- TITRAMAX 100 / 101 / 1000 (60 pages)
- UNIMAX 1010 (68 pages)
- UNIMAX 2010 (72 pages)
- Incubator 1000 (64 pages)
- Pump drives (198 pages)
- SilentCrusher M (146 pages)
- SilentCrusher S (98 pages)