110-509 HD/EN

20

Setting the High Limit Temperature

This sub-menu allows you to limit how high the temperature set point may be set. It also serves as a high
limit safety, alerting you if bath temperature rises above the high limit temperature setting. The High Limit
value may be set from +40° to +155°C / +100° to +305°F; the factory default is +150°C / +300°F.

To avoid an unwanted shutdown during regular operation, the High Limit value should be set at least 5°
higher than the selected control temperature.

To Access:

Press the key until HIGH LIMIT appears on the display.

To Change:

Press or .

To Accept:

Press or allow the LCD to return to the main operational display

(approximately 10 seconds).


Appears when in the
High Limit sub-menu