Routine maintenance and troubleshooting – Heidolph RotaChill User Manual

Page 26

background image

110-509 HD/EN

25

Routine Maintenance and Troubleshooting

WARNING:

Always turn your Circulating Bath OFF and disconnect it from the electrical power outlet

before performing any maintenance or service.

WARNING:

To avoid the potential for burns, allow the Circulating Bath to cool completely before

cleaning or performing any maintenance.

WARNING:

Always drain all fluid from the reservoir before moving or lifting your Circulator. Be sure

to follow your organization’s procedures and practices regarding the safe lifting and relocation of
heavy objects.

Maintaining Clear Bath Water

Optimum temperature and moisture conditions for algae growth exist when using water as a bath fluid.

WARNING:

Do not use chlorine bleach.

Draining the Bath Reservoir

WARNING:

Bath fluids should be stored and disposed of according to applicable laws and

regulations.

Refrigerated / Heating and Heat only Circulating Baths

Refrigerated / Heating and Heat only Circulating Baths are equipped with a drain valve and port located
either behind the front access panel or on the right hand side of the unit.

To drain fluid from the bath, attach a short length of suitable 11.5 mm ID / 0.45 inch ID tubing to the drain
port and secure it using a hose clamp with a minimum ID of 18 mm / 0.7 inch. Open the drain valve using
a flat blade screwdriver. When closing the valve, do not over tighten.

WARNING:

Be sure to close the drain valve before refilling the bath reservoir. Do not over tighten.

Drain Port

Drain Valve