Heidolph RotaChill User Manual
Page 17

110-509 HD/EN
16
Set-Up Sub-Menus
Press the key to access the Temperature Controller’s set-up sub-menus. The or keys are used
to change the current setting / value in the sub-menu.
Sub-Menu
Selection / Range
To Change
Current Value
Temperature unit
°C or °F
Press for °C
Press for °F
Offset Calibration
-3.0°C to +3.0°C
Press or
Low Limit
-55° to +10°C / -65° to +50°F
Press or
High Limit
+40° to +155°C / +100 to +305°F
Press or
Auto Cool
1° to 135°C / 34° to 275°F
Press or
To accept a value in a sub-menu, press or allow the display to ‘time out’ (approximately 10 seconds).
Adjusting the Temperature Set Point
This is the temperature at which the fluid in your Circulating Bath will be maintained. It may be set to one-
tenth of a degree over a range of -50.0° to +135.0°C / -60.0° to +275.0°F. The factory default set point is
+20.0°C / +68.0°F.
To Change:
Press the or key. The word “SET” will begin flashing. Continue holding
or repeatedly press or until the desired set point temperature is displayed.
To Accept:
Press or allow the LCD to return to the main operational display
(approximately 10 seconds).
Flashes when set
point is being
changed