Key-in span calibration value


Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid span calibration value, obtained from a successful H17 calibration
procedure, must be used.

1. While in the Setup mode, scroll to "H 20", then scroll down once using the ZERO key.
2. The display will momentarily show "CAL 1", followed by a flashing zero. Use the four

directional keys (shown in Figure 6-1) to adjust the displayed value to the span
calibration value.

3. After setting the exact value, press the NET/GROSS key to save the value. If H15=2 or

H15=3, the display will also show “CAL 2” or “CAL 3”, the two parameters can be set
too.

4. If the entered value is greater than zero, the display will show "E CAL 1" momentarily

then revert back up to H20. If a value of zero is entered, the indicator will briefly
show "Err 5" then revert back to the screen described above in Step # 2.