NAME/CODE
DESCRIPTION
CODE/VALUE
H1
Graduation
Specifies number of full-scale graduations.
Value should be consistent with legal
requirements and environmental limits on the
useful system resolution.
500
1,000 1,500
2,000 2,500
3,000 4,000
5,000√ 6,000
8,000 10,000
12,000 20,000
30,000 40,000
50,000
H3
Zero Track
Band
Selects the range within which the scale will
automatically zero. Note that the scale must be
in standstill to automatically zero. Selections
are in Display Divisions.
0d
0.5d √
1d
3d
5d
H4
Zero Range
Selects the range which the scale may be
zeroed. Note that the indicator must be in
standstill to zero the scale.
100% √
1.90%
H5
Motion Band
Sets the level at which motion is detected by
comparing the present display update with the
previous one. If motion is not detected for two
seconds or more, scale is in standstill and can
process a Print or Zero command. Maximum
value varies depending on local regulations
o.25d
1d
3d√
5d
10d
H6
Digital Filter
Averages weight readings to produce higher
stability. The higher the filter setting, the
stability but the slower the indicator's response
time. Choose SLO unless a very fast response
is needed.
Fast
Nned√
slo
H7
Overload
limit
Selects the desire formula which determines
the point at which the indicator shows overload.
All selections are based on the primary unit
selected in H8. "FS"=Full
scale in primary units.
FS
FS+2%√
FS+1d
FS+9d
H8
Calib. Unit
Selects the primary base unit to be used in the
calibration process. Also the default unit for
normal operation.
1√
2
"1"=primary unit is lb. "2"=primary unit is kg