DIGIWEIGH DWP-L Series User Manual

Page 11

background image

10

NAME/CODE

DESCRIPTION

CODE/VALUE

H1
Graduation

Specifies number of full-scale graduations.

Value should be consistent with legal

requirements and environmental limits on the

useful system resolution.

500
1,000 1,500
2,000 2,500
3,000 4,000
5,000√ 6,000
8,000 10,000
12,000 20,000
30,000 40,000
50,000

H3
Zero Track
Band

Selects the range within which the scale will

automatically zero. Note that the scale must be

in standstill to automatically zero. Selections

are in Display Divisions.

0d
0.5d
1d
3d

5d

H4
Zero Range

Selects the range which the scale may be

zeroed. Note that the indicator must be in

standstill to zero the scale.

100%
1.90%

H5
Motion Band

Sets the level at which motion is detected by

comparing the present display update with the

previous one. If motion is not detected for two

seconds or more, scale is in standstill and can

process a Print or Zero command. Maximum

value varies depending on local regulations

o.25d

1d

3d√

5d
10d

H6
Digital Filter

Averages weight readings to produce higher

stability. The higher the filter setting, the

stability but the slower the indicator's response

time. Choose SLO unless a very fast response

is needed.

Fast

Nned√

slo

H7
Overload
limit

Selects the desire formula which determines

the point at which the indicator shows overload.

All selections are based on the primary unit

selected in H8. "FS"=Full

scale in primary units.

FS
FS+2%√
FS+1d
FS+9d

H8
Calib. Unit

Selects the primary base unit to be used in the

calibration process. Also the default unit for

normal operation.

1√
2

"1"=primary unit is lb. "2"=primary unit is kg