DIGIWEIGH DWP-L Series User Manual

Page 12

background image

11

NAME/CODE

DESCRIPTION

CODE/VALUE

H9

Display

Divisions

Determines the desired weight increments.

Value should be consistent with legal

requirements.

1√

2

3

H10

Decimal Pt.

Determines location of the decimal point.

0 √ 0.0
0.00 0.000
0.0000 00

H15

Span Calib.

Type

Selects the Span Calibration type. 1 -one time

calib in the graduation. 2 - two times calib in the

graduation. 3 - three times calib in the

graduation.

1√
2
3

H16

Zero

Calibration

Places indicator into the zero calibration

routine. Scrolling down with the ZERO key one

level begins the procedure.

Press ZERO key to

begin sequence

H17

Span

Calibration

Places indicator into the span calibration

routine. Scrolling down with the ZERO key one

level begins the procedure.

Press ZERO key to

begin sequence

H18

View

Calibration

Actuates the function that allows you to view

both the zero and span calibration value. The

values displayed in this function are valid only

after Calibration (H16 & H17) has been

successfully completed. Scrolling down with

the ZERO key one level begins the procedure.

Press ZERO key to

begin sequence

H19

Key-in Zero

Allows you to key-in zero calibration value in

case of memory loss in the field. Scrolling

down with the ZERO key one level begins the

procedure.

Press ZERO key to

begin sequence

H20

Key-in Span

Allows you to key-in a known span calibration

value in case of memory loss in the field.

Scrolling down with the ZERO key one level

begins the procedure.

Press ZERO key to

begin sequence

H21

Factory Reset

This sub-menu will reset parameters in the "H"

and "A" menu to the default settings. USE WITH

CAUTION!

Press ZERO key to

begin sequence