DIGIWEIGH DWP-102E User Manual
Page 9
- -
7
KEY 5
To set the MOD Key for animal weighing parameters setting.
“F4 CHK”: Check-weighing setting
Options: BEEP 1/2/3.
“F5 Brt”: Adjust the brightness of LED display. For LED version
Select 1-8 : 8 is most bright
“F5 INP”: Display the internal counts For LCD version .
Used to check the connection of loadcell and working status of ADC circuit.
Press “TARE” to exit.
“F6 INP”: Display the internal counts For LED version
“F6 CAL”: Calibration for LCD Version. See below:
“F7 CAL”: Calibration
“PN”: Press FUN Key to enter “PN”. Initiate password is 0000
“P1 SP”:
Set AD converting speed. Options: 1/2/3/4/5. 1 is the fastest and 5 is the
slowest.
“P2 AUT”: Select the “Power on “ zero maintenance, Options: 0/2/5/10/20/50/100% of
the capacity.
“P3 RAN”: Select the manually zero maintenance, Options: 2/ 4/10/20/50/100% of the
capacity.
“P4 AZ”: Select the auto zero maintenance, Options: 0.5d/1d/2d/4d. “d” means division.
“P5 DEC”: Set the decimal point. Options: 0/0.0/0.00/0.000/0.0000.
“P6 CAP”: Set the capacity.
“P7 INC”: Set division increment. Options: 1/2/5/10/20/50.
“P8 CAL”: Calibration
After finishing all necessary settings above, the indicator is ready for calibration. (Note: If
the scale has settings already done you can directly calibrate it.) Press FUN Key when
displaying
“P8 CAL”. Display will show “noLoAd”. Empty the platform by removing the
weight if any. When platform is empty and STABLE indicator is on then press FUN Key to
confirm. Display shows
“XXXXXX”. Set it to be the weight of the weigher you want to put
on. Press FUN Key to confirm, and it
will show “LoAd”. Put the calibration weigher on the
platform and press FUN Key after the STABLE indicator is on. Indicator will go to weighing
state. The calibration is finished.
“P9 INT”: Initialize the memory to ex-work state.
“P10 RS”: Communication mode.
This option is used for the RS232C serial port m
ode setup. Options: “RS Cot” is
continuous Data Transmission for using Remote Display o
r External Display, “RS Prt” is
used for the Serial Printer.
“RS Aut” is used for “transmission after every weighing”.
“P11 FL”: Weight adjustment for acceleration of gravity.