beautypg.com

DIGIWEIGH DWP-102E User Manual

Page 3

background image

- -

1

SECTION 1 INTRODUCTIONS

1.1 Introduction

The indicator provides an accurate, fast and versatile series of general purpose weighing

indicator with counting, check-weighing, animal weighing functions and preset tare.

1.2 Specification

Housing

plastic housing

Power supply

110V 50/60Hz

built-in rechargeable battery. 6V4AH.

Display

6 digits 1

’LED.

Max input

15mV

ADC

Σ-Δ

External Resolution

1/30000

Interface

RS-232 Output

Stabilisation time

1 second typical

Operating temperature

0°C - 50°C / 32°F - 104°F

Load cell drive voltage

5V/ Max 150mA

Other function

Real-time clock and units conversion

Load cells

Up to eight 350 ohms cells

1.3 Installation

The weighing indicator should be sited in a location accuracy would not be degraded.

Avoid extremes of temperature.

Avoid unstable power sources. Do not use near large users of electricity such as welding

equipment or large motors.

Verify the voltage showing on the label matches the voltage in your area.

connection of loadcell:

Plastic housing connect to Din9

Pin6 +E

Pin1 -E

Pin9 +S

Pin8 -S

Pin5 shield

connection of RS232

Plastic housing connect to Din9 Pin 2 TXD Pin3 RXD Pin5 GND

SECTION 2 KEY DESCRIPTIONS

SOFT POWER ON/OFF Key