Carbolite 3216P1/P5 User Manual
Page 11

Programming
11
MC18-GB
– v1.04
Ramp Units
Press scroll
until the display shows
RAMP.U <Ramp Units>: Use the Up
Down
to
select the Ramp Units of Hour, Min or Seconds.
Dwell Units
Press Scroll
until the display shows
DWEL.U <Dwell Units>: Use the Up
Down
to
select the Dwell Units of Hour or Min.
Holdback (See section 4.7)
Press Scroll
until the display shows
H.BACK <Program Hold Back>: If a Holdback value is
required, use the Up
Down
to enter the required value or select “OFF” if no holdback is
required.
Ramp Rate
Press Scroll
until the display shows
RMP.1 <Ramp Rate 1>: Using the Up
Down
enter the
value for the first ramp rate depending on the Ramp Units selected. If the Ramp
segment is not required select “OFF”.
Target Setpoint
Press Scroll
until the display shows
T.SP 1 <Target SP 1>: Enter the temperature that
you want the unit to ramp up to using “Ramp rate 1”.
If “Ramp Rate 1” has been set to “OFF”. The unit will Heat/Cool directly to the Target SP.
Dwell Time
Press Scroll
until the display shows
DWELl. 1 <DWELL TIME 1>: Enter the time to dwell
at “Target SP 1”. If the dwell segment is not required, use the Down
to select
OFF, which
is below the Zero value.
This Process is repeated for each of the 8 segments of the program.
If not all the segments are used for a program, the Ramp & Dwell of each of the subsequent
Segments should be set to
OFF.
A program will end in one of two ways, either revert to the control Set Point or dwell at the
temperature set in the last segment used. When a program finishes on a dwell and the dwell
time expires the temperature will revert to the control setpoint.
Note: Before running a program ensure that the control set point is set to Zero to avoid
unexpected heating at the end of the program.