Fitting the optional insulating plugs and radiation shield

For optimum temperature uniformity insulating plugs or radiation shields should be placed in the
work tube ends as shown in

Figure 2-21

,

Figure 2-20

and

Figure 2-22

. With an extended work tube

the stem (8) of the insulation plug assembly lines up with the end of the work tube.

If a gas/vacuum end seal is fitted (4), either horizontally or vertically, the insulation plug hooks
onto the end seal hook; see

Figure 2-20

.

Alignment of radiation shields is similar to that of insulation plugs see

Figure 2-22

.

If a metal work tube is being used in the furnace, ensure that it is earthed. See the safety warning
in section 3.5.

Index to numbered items in

Figure 2-21

,

Figure 2-20

and

Figure 2-22

1. Extended work tube guard
2. Standard work tube
3. Extended work tube
4. Gas/vacuum end seal
5. Insulation plug for standard work tube
6. Insulation plug for extended work tube

7.

Radiation shield for extended work tube

8.

Insulation plug stem

Figure 2-20. Cross-section view showing extended work tube
insulation plug fitted.

Figure 2-21. Cross-section view showing
standard length work tube insulation plug
fitted.

Figure 2-22. Cross-section view showing extended work tube
radiation shield fitted