Carbolite TZF Series User Manual
Page 15
MF12 3.12
15
The models covered by this manual are designed to run on 100% power, with the exceptions
below.
Model
Volts:
200V
208V
220V
230V
240V
380V
400V
415V
Equal zone models:
TZF 12/38/400E
TZF 12/65/550E
100
100
89
82
75
-
-
-
Custom Models:
TZF 12/25/550
TZF 12/38/550
TZF 12/75/550
TZF 12/75/900
100
100
89
82
75
-
-
-
Standard Model:
TZF 12/100/900
(from April 2000)
100
100
89
82
75
89
82
75
Other possible exceptions:
Further custom furnaces based on the standard models may be designed with power limits other
than 100%.
Furnaces supplied for use on 254 Volt typically have a power setting of 89% of the limit for the
corresponding 240V model.
See also other documents in the category Carbolite Kiln:
- Peak Series (16 pages)
- AX30 (12 pages)
- GP Series (16 pages)
- GPC Series (16 pages)
- LHT Series (16 pages)
- HRF (MF05) (16 pages)
- HRF 7/45 (12 pages)
- ABA 7/35B (72 pages)
- TLD Series (12 pages)
- CR Series (16 pages)
- MTT 12/38/850 (16 pages)
- CF 15 - CF 60 (20 pages)
- PTC 12/20/150 (16 pages)
- CFM Series (16 pages)
- AAF Series (16 pages)
- SNF Series (12 pages)
- HZS Series (16 pages)
- Work Tube End Seals (12 pages)
- MTF Series (16 pages)
- TZF Series (16 pages)
- GHA Series (24 pages)
- STF Series (20 pages)
- EHA Series (28 pages)
- CWF Series (16 pages)
- ABF 8/28 (16 pages)
- VCF Series (16 pages)
- ELF Series (16 pages)
- HTR 11/75 (16 pages)
- HST Series (16 pages)
- STF Series (20 pages)
- VST 17/-- (16 pages)
- BLF 17/3 (16 pages)
- RHF Series (17 pages)
- HTF 17 (20 pages)
- RHF Series 3 litre - 35 litre (16 pages)
- CDF 15/1b (16 pages)
- CAF 16/38 (16 pages)
- CAF DIGITAL IMAGING (52 pages)