beautypg.com

Specifications – Carbolite GVC Series User Manual

Page 21

background image


MF48 -1.18

21

9.0 S

PECIFICATIONS

9.1 Models Covered by this Manual

GHA models – horizontal single-zone tube furnaces with controls in integral base
GVA models – versatile single-zone tube furnaces with controls in integral base
GHC models – horizontal three-zone tube furnaces with controls in integral base
GVC models – versatile three-zone tube furnaces with controls in integral base
All models have cylindrical elements with wire formed in the surface of the insulation material.
All models have can accept work tubes up to a maximum outside diameter of 170mm.
All models have a maximum operating temperature of 1200°C (1100° continuous).
Model names are of the type “GHA 12/300” where 300 is the heated length. Sizes range from
300mm heated length (GHA/GVA) or 450mm (GHC/GVC) to 1200mm (all models).

Heated
Length

Max.

Power

Minimum
work tube

length

Type C

work tube

Type D

work tube

(mm)

(W)

(mm)

(mm)

(mm)

300

2340

480

600

900

450

3120

630

750

1050

500

3900

780

900

1200

750

4680

930

1050

1350

900

5460

1080

1200

1500

1050

6240

1230

1350

1650

1200

7020

1380

1500

1800

1500

8580

1530

1800

2100

Work tubes may be supplied in any commercially available size up to the maximum of 170mm
o/d. Type C protrudes about 40mm from each end of the furnace (from the end guards), while type
D allows for the fitting of gas seals or other attachments.

9.2 Environment

The furnaces contain electrical parts and should be stored and used in indoor conditions as
follows:

temperature:

5°C - 40°C

relative humidity:

maximum 80% up to 31°C decreasing linearly to 50% at 40°C

Carbolite reserves the right to change specifications without notice.