Carbolite VMF Series User Manual
Page 4
AAF, GSM, VMF
4
MF06 – 3.32
L3
Black
to phase 3 except 2-phase
N
Light Blue
to neutral except delta
PE
Green/Yellow
to earth (ground)
2.4 “Universal Wiring”
Year 2000: 18 litre 3-phase models only
Year 2001 onwards: 18 litre 1-phase and 3-phase
The AAF 11/18 and 12/18 can be easily rewired between 1-phase and 3-phase supplies. This
applies to 3-phase+N and 3-phase delta in the ranges 380/220V–415/240V and 208-240V, but
does not apply to 3-phase star without neutral (e.g. 380V).
To alter the configuration, remove the back panel and alter the wiring connections between the
supply terminal block and the EMC filters, using the appropriate diagram from section 7.4.
Models made for single phase only (e.g. before year 2001) are not affected and cannot be
converted to 3-phase.
If in doubt, please consult Carbolite.